Benzene and VCM traces in drinking water
Column: MEGA-VOC 1 – 0.25mm, 1.40µm, 50m
Catalog Code: C-VOC1-025-140-50
USP G43 stationary phase
Conditions:
njection: Splitless (1min), 300°C.
Detector: FID, 260°C.
Oven Program: 70°C (4min), 20°C/min, 160°C (3min). Carrier Gas: Helium, 1mL/min.
Sampling: 10mL of drinking water + 25% of NaCl.
Extraction: HS-SPME (CAR-PDMS 85 μm fiber, 2 min of Magnetic stirring, 20min of fiber exposition time @ 50°C, 4min desorption time)
Sample:
Natural water added with a standard mix, VCM (0,1 μg/L) Benzene (0,04 μg/L).